PHI 5600: COOLING SYSTEM

Description of cooling system,

  • Heat Exchange system circulates water through both sources and it has a radiator to cool down the water. 
  • Besides the water reservoir it has a water filter and a de-ionizer cartridge which operates continuously. 
  • It also has a flow switch to prevent the X-ray source from turning on if there is now flow of water (without cooling the anode would burn very quickly)
  • There is also electronic control of the pump, (see pictures for position of cable for pump control). The cable is labelled A179J5 and A179.
  • Uses distilled water (3 gallons)
  • The direction of flow is shown in the pictures.
  • It gets connected to 208V power strip.
  • A description of the production of x-rays is shown in the video.

 

 

     