The PHI 5600 is typically on 24 hours a day and not powered on and off daily. The system contains a circuit breaker panel, a power control panel, and the emergency power off switch. The individual circuit breakers and power control panel allow individual segments of the instrument console to be turned on and off for servicing purposes. In the event of a hazardous occurrence or emergency, the entire system can be shut off by depressing the Emergency Off button.
In the event of a power interruption, the system will remain in an off state and wait for you to restore power. The following steps should be performed to restore power:
Press the water pump start switch on the front panel of the 50-096 X-ray source control to start the water pump for the X-ray source.
Log onto the computer and start the PHI PC-ACCESS and MultiPak software packages
Several electron/x-ray optics devices in the 5800 systems use hot filaments: the dual anode x-ray source, the monochromator x-ray source, the ion gun, and the electron neutralizer. The x-ray sources are typically run in an automated mode and are shut off at the end of an analysis sequence. The ion gun and electron neutralizer typically idle in a standby mode, where the filament is on, but accelerating voltages are off. At the end of the day (if your analysis is complete), these devices should be turned off by selecting the off buttons in the ion gun and neutralizer software menus.